Cardiac output measurements may inform diagnosis and provide guidance of therapeutic interventions in patients with hemodynamic instability. The FloTrac™ algorithm uses uncalibrated arterial pressure waveform analysis to estimate cardiac output. Recently, a new version of the algorithm has been developed. The aim was to assess the agreement between FloTrac™ and routinely performed cardiac output measurements obtained by critical care ultrasonography in patients with circulatory shock. A prospective observational study was performed in a tertiary hospital from June 2016 to January 2017. Adult critically ill patients with circulatory shock were eligible for inclusion. Cardiac output was measured simultaneously using FloTrac™ with a fourth-generation algorithm (CO Eighty-nine paired cardiac output measurements were performed in 17 patients during their first 24 h of admittance. CO In critically ill patients with circulatory shock, there was disagreement and clinically unacceptable trending ability between values of cardiac output obtained by uncalibrated arterial pressure waveform analysis and critical care ultrasonography. Clinicaltrials.gov, NCT02912624, registered on September 23, 2016.
Disagreement in cardiac output measurements between fourth-generation FloTrac and critical care ultrasonography in patients with circulatory shock: a prospective observational study
T. Kaufmann, R. Clement, B. Hiemstra, J. Vos, T. Scheeren, F. Keus and I. van der Horst
J Intensive Care 2019;7:21.