To evaluate the feasibility of obtaining hemodynamic metrics of echocardiographically derived 3-dimensional printed mitral valve models deployed in a pulse-duplicator chamber. Exploratory study. Tertiary-care university hospital. Percutaneous MitraClip procedure patient. Three-dimensional R-wave gated, full-volume transesophageal echocardiography images were obtained after deployment of the MitraClip device. A high-quality diastolic frame of the mitral valve was segmented using Mimics Innovation Suite and merged with a flange. The data were exported as a stereolithography (.stl) file, and a rigid 3-dimensional model was printed using a MakerBot Replicator 2 printer. A flexible silicone cast then was created and deployed in the pulse-duplicator chamber filled with a blood-mimicking fluid. The authors were able to obtain continuous-wave Doppler tracings of the valve inflow with a transesophageal echocardiography transducer. They also were able to generate diastolic ventricular and atrial pressure tracings. Pressure half-time and mitral valve area were computed from these measurements. This pulse duplicator shows promising applications in hemodynamic testing of patient-specific anatomy. Future modifications to the system may allow for visualization and data collection of gradients across the aortic valve.
Hemodynamic Testing of Patient-Specific Mitral Valves Using a Pulse Duplicator: A Clinical Application of Three-Dimensional Printing
A. Mashari, Z. Knio, J. Jeganathan, M. Montealegre-Gallegos, L. Yeh, Y. Amador, R. Matyal, R. Saraf, K. Khabbaz and F. Mahmood
J Cardiothorac Vasc Anesth 2016;30(5):1278-85.